CFR 1020.32 Fluoroscopic equipment (f) Indication of potential and current
Jump to navigation
Jump to search
(f) Indication of potential and current During fluoroscopy and cinefluorography, x-ray tube potential and current shall be continuously indicated. Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).