CFR 1020.32 Fluoroscopic equipment (b)(5) Rectangular image receptors

From knowledgebase
Jump to navigation Jump to search

5) Fluoroscopy and radiography using the fluoroscopic imaging assembly with inherently rectangular image receptors.

For x-ray systems manufactured on or after June 10, 2006, the following applies:

(i) Neither the length nor the width of the x-ray field in the plane of the image receptor shall exceed that of the visible area of the image receptor by more than 3 percent of the SID. The sum of the excess length and the excess width shall be no greater than 4 percent of the SID.

(ii) The error in alignment shall be determined along the length and width dimensions of the x-ray field which pass through the center of the visible area of the image receptor.