CFR 1020.32 Fluoroscopic equipment (b)(3) Non-image-intensified fluoroscopy
Jump to navigation
Jump to search
(3) Non-image-intensified fluoroscopy The x-ray field produced by non-image-intensified fluoroscopic equipment shall not extend beyond the entire visible area of the image receptor. Means shall be provided for stepless adjustment of field size. The minimum field size, at the greatest SID, shall be containable in a square of 5 cm by 5 cm.