CFR 1020.32 Fluoroscopic equipment (d)(3) Measuring compliance

From knowledgebase
Revision as of 07:20, 25 August 2020 by Dirkjan (talk | contribs) (Created page with "''(3) Measuring compliance'' Compliance with paragraph (d) of this section shall be determined as follows: * (i) If the source is below the x-ray table, the AKR shall be measu...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

(3) Measuring compliance Compliance with paragraph (d) of this section shall be determined as follows:

  • (i) If the source is below the x-ray table, the AKR shall be measured at 1 cm above the tabletop or cradle.
  • (ii) If the source is above the x-ray table, the AKR shall be measured at 30 cm above the tabletop with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement.
  • (iii) In a C-arm type of fluoroscope, the AKR shall be measured at 30 cm from the input surface of the fluoroscopic imaging assembly, with the source positioned at any available SID, provided that the end of the beam-limiting device or spacer is no closer than 30 cm from the input surface of the fluoroscopic imaging assembly.
  • (iv) In a C-arm type of fluoroscope having an SID less than 45 cm, the AKR shall be measured at the minimum SSD.
  • (v) In a lateral type of fluoroscope, the air kerma rate shall be measured at a point 15 cm from the centerline of the x-ray table and in the direction of the x-ray source with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement. If the tabletop is movable, it shall be positioned as closely as possible to the lateral x-ray source, with the end of the beam-limiting device or spacer no closer than 15 cm to the centerline of the x-ray table.