CFR 1020.32 Fluoroscopic equipment (b)(6) Override capability

From knowledgebase
Revision as of 07:20, 25 August 2020 by Dirkjan (talk | contribs) (Created page with "''(6) Override capability'' If the fluoroscopic x-ray field size is adjusted automatically as the SID or image receptor size is changed, a capability may be provided f...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

(6) Override capability If the fluoroscopic x-ray field size is adjusted automatically as the SID or image receptor size is changed, a capability may be provided for overriding the automatic adjustment in case of system failure.

If it is so provided, a signal visible at the fluoroscopist's position shall indicate whenever the automatic field adjustment is overridden. Each such system failure override switch shall be clearly labeled as follows:

For X-ray Field Limitation System Failure