CFR 1020.32 Fluoroscopic equipment (b)(3) Non-image-intensified fluoroscopy

From knowledgebase
Revision as of 07:20, 25 August 2020 by Dirkjan (talk | contribs) (Created page with "''(3) Non-image-intensified fluoroscopy'' The x-ray field produced by non-image-intensified fluoroscopic equipment shall not extend beyond the entire visible area of the image...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

(3) Non-image-intensified fluoroscopy The x-ray field produced by non-image-intensified fluoroscopic equipment shall not extend beyond the entire visible area of the image receptor. Means shall be provided for stepless adjustment of field size. The minimum field size, at the greatest SID, shall be containable in a square of 5 cm by 5 cm.